Transcatheter implantation of ballon-expandable valves with ballon underexpansion in patients with borderline aortic valve annulus

Abstract

Background. Recently transcatheter aortic valve implantation (TAVI) has become the optimal choice procedure for patients with aortic stenosis. Last generation of transcatheter valves allows to perform TAVI in a vast majority of patients. Patients with borderline aortic annulus size who have transition of aortic ring parameters (diameter, area and perimeter of ring) from a smaller to a larger size comprise the particular category of patients.

Aim – analysis of TAVI results with use of underexpanded ballon-expandable prosthesis in patients with borderline aortic valve annulus was the aim of this paper.

Material and methods. Among 310 patients who underwent TAVI in 2019 at cardiovascular department of National Cardiology Research Center 38 (12%) patients had borderline aortic valve annulus and they were treated with an underexpansion of ballon-expandable prosthesis approach. Clinico-pathologic profile and hospital outcomes were analyzed.

Results. Mean values of maximal and mean pressure gradient were 88.2±28.3 and 54.4±19.1 mm Hg respectively. Mean area of aortic valve ring was 479.1±88.2 mm2. Mean reduction of volume accounted for 1.9; 2.1 and 2.4 for 23, 26 and 29 sizes respectively. Technical success was 100%. There were no high pressure gradients at 5 day postoperatively with mean values of maximal and mean pressure gradient 19±7,0 and 10±4,0 mm Hg respectively. Complications included 1 major vascular complication (endarterectomy from vascular access site) and 1 case of pacemaker implantation due to atrioventricular block.

Conclusion. Underexpansion of ballon-expandable prosthesis is an effective approach in patients with borderline aortic ring size.

Keywords:transcatheter aortic valve implantation, ballon- expandable prosthesis, “valve-in-valve”

Conflict of interest. The authors declare no conflict of interest.
For citation: Salichkin D.V., Imaev T.E., Komlev A.E., Lepilin P.M., Kolegaev A.S., Makeev M.I., Akchurin R.S. Transcatheter implantation of ballon-expandable valves with ballon underexpansion in patients with borderline aortic valve annulus. Clin Experiment Surg. Petrovsky J. 2020; 8 (2): 17–24. DOI: 10.33029/2308-1198-2020-8-2-17-24 (in Russian)
Received 10.12.2019. Accepted 26.03.2020.

References

1. Cribier A., Eltchaninoff H., Bash A., Borenstein N., Tron C., Bauer F., et al. Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description. Circulation. 2002; 106: 3006-8.

2. Vahanian A., Alfieri O., Andreotti F., Antunes M.J., Baron-Esquivias G., Baumgartner H., et al. Guidelines on the management of valvular heart disease (version 2012): the joint task force on the management of valvular heart disease of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS). Eur J Cardiothorac Surg. 2012; 42: S1-44.

3. Holmes D.R. Jr., Mack M.J., Kaul S., Agnihotri A., Alexander K.P., Bailey S.R., et al. 2012 ACCF/AATS/SCAI/ STS expert consensus document on transcatheter aortic valve replacement: developed in collaboration with the American Heart Association, American Society of Echocardiography, European Association for Cardio-Thoracic Surgery, Heart Failure Society of America, Mended Hearts, Society of Cardiovascular Anesthesiologists, Society of Cardiovascular Computed Tomography, and Society for Cardiovascular Magnetic Resonance. Catheter Cardiovasc Interv. 2012; 79: 1023-82.

4. Siddique S., Gada H., Mumtaz M.A., Vora A.N. Should All Low-risk Patients Now Be Considered for TAVR? Operative Risk, Clinical, and Anatomic Considerations. Curr Cardiol Rep. 2019; 21 (12): 161.

5. Sousa O., Ponte M., Caeiro D., Carvalho M., Leite D., Rocha J., et al. Transcatheter aortic valve implantation: Is anatomy still the limiting factor? Rev Port Cardiol. 2013; 32 (4): 281-6.

6. Willson A.B., Webb J.G., Labounty T.M., et al. 3-dimensional aortic annular assessment by multidetector computed tomography predicts moderate or severe paravalvular regurgitation after transcatheter aortic valve replacement: a multicenter retrospective analysis. J Am Coll Cardiol, 2012; 59: 1287-94.

7. Barbanti M., Yang T.H., Rodés-Cabau J., et al. Anatomical and procedural features associated with aortic root rupture during balloon-expandable transcatheter aortic valve replacement. Circulation. 2013; 128: 244-53.

8. Binder R.K., Webb J.G., Willson A.B., et al. The impact of integration of a multidetector computed tomography annulus area sizing algorithm on outcomes of transcatheter aortic valve replacement: a prospective, multicenter, controlled trial. J Am Coll Cardiol. 2013; 62: 431-8.

9. Barbanti M., Leipsic J., Binder R., Dvir D., Tan J., Freeman M., et al. Underexpansion and Ad Hoc Post-Dilation in Selected Patients Undergoing Balloon-Expandable Transcatheter Aortic Valve Replacement. JACC. 2014; 63 (10): 976-81.

10. Kasel A.M., Cassese S., Bleiziffer S., Amaki M., Hahn R.T., Kastrati A., et al. Standardized imaging for aortic annular sizing: implications for transcatheter valve selection. JACC Cardiovasc Imaging. 2013; 6 (2): 249-62.

11. Barbanti M., Leipsic J., Binder R., Dvir D., Tan J., Freeman M., et al. Underexpansion and Ad Hoc Post-Dilation in Selected Patients Undergoing Balloon-Expandable Transcatheter Aortic Valve Replacement. J Am Coll Cardiol. 2014; 63 (10): 976-81.

12. Willson A.B., Webb J.G., Labounty T.M., et al. 3-dimensional aortic annular assessment by multidetec- tor computed tomography predicts moderate or severe paravalvular regurgitation after transcatheter aortic valve replacement: a multicenter retrospective analysis. J Am Coll Cardiol. 2012; 59: 1287-94.

13. Binder R.K., Webb J.G., Willson A.B., et al. The impact of integration of a multidetector computed tomography annulus area sizing algorithm on outcomes of transcatheter aortic valve replacement: a prospective, multicenter, controlled trial. J Am Coll Cardiol. 2013; 62: 431-8.

14. Barbanti M., Yang T.H., Rodés-Cabau J., et al. Anatomical and procedural features associated with aortic root rupture during balloon-expandable transcatheter aortic valve replacement. Circulation. 2013; 128: 244-53.

15. Martin C., Sun W. Transcatheter valve underex-pansion limits leaflet durability: implications for valve-in-valve procedures. Ann Biomed Eng. 2017; 45 (2): 394- 404. DOI: https://doi.org/10.1007/s10439-016-1738-8.

16. Chen H.L., Liu K. Clinical outcomes for trans- catheter valve-in-valve in treating surgical bioprosthetic dysfunction: A meta-analysis. Int J Cardiol. 2016; 212: 138-141.

17. Webb J.G., Dvir D. Transcatheter aortic valve replacement for bioprosthetic aortic valve failure: the valve- in-valve procedure. Circulation. 2013; 127: 2542-50.

18. Dvir D., Webb J.G., Bleiziffer S., Pasic M., Waks- man R., Kodali S., et al. Valve-in-Valve International Data Registry. Transcatheter aortic valve implantation in failed bioprosthetic surgical valves. JAMA. 2014; 312: 162-70.

19. Harbaoui B., Courand P.Y., Schmitt Z., Farhat F., Dauphin R., Lantelme P. Early Edwards SAPIEN Valve Degeneration After Transcatheter Aortic Valve Replacement. JACC Cardiovasc Interv. 2016; 9: 198-9.

20. Cribier A., Eltchaninoff H., Bash A., Borenstein N., Tron C., Bauer F., et al. Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description. Circulation. 2002; 106: 3006-8.

21. Schultz C.J., Weustink A., Piazza N., Otten A., Mollet N., Krestin G., et al. Geometry and Degree of Apposition of the CoreValve ReValving System With Multislice Computed Tomography After Implantation in Patients With Aortic Stenosis. J Am Coll Cardiol. 2009; 54: 911-8. 

22. Willson A.B., Webb J.G., Gurvitch R., Wood D.A., Toggweiler S., Binder R., et al. Structural integrity of balloon-expandable stents after transcatheter aortic valve replacement: assessment by multidetector computed tomography. JACC Cardiovasc Interv. 2012; 5: 525-32. 

23. Shalabi A., Spiegelstein D., Sternik L., Feinberg M.S., Kogan A., Levin S., et al. Sutureless Versus Stented Valve in Aortic Valve Replacement in Patients With Small Annulus. Ann Thorac Surg. 2016; 102: 118-22.

24. Milburn K., Bapat V., Thomas M. Valve-in-valve implantations: is this the new standard for degenerated bioprostheses? Review of the literature. Clin Res Cardiol. 2014: 1-13.

25. Allen K.B., Chhatriwalla A.K., Cohen D.J., Saxon J.T., Aggarwal S., Hart A., et al. Bioprosthetic Valve Fracture to Facilitate Transcatheter Valve-in-Valve Implantation. Ann Thorac Surg. 2017; 104: 1501-8.

All articles in our journal are distributed under the Creative Commons Attribution 4.0 International License (CC BY 4.0 license)

CHIEF EDITOR
CHIEF EDITOR
Sergey L. Dzemeshkevich
MD, Professor (Moscow, Russia)

Journals of «GEOTAR-Media»